„Apera Instruments AI532“
Apera Instruments EC820 Benchtop Conductivity Meter Kit
Vartotojo vadovas
1. Įvadas ir produkto apžvalgaview
The Apera Instruments EC820 Benchtop Conductivity Meter is a high-precision instrument designed for accurate measurement of Conductivity (EC), Total Dissolved Solids (TDS), Salinity, Resistivity, and temperature. This meter is equipped with advanced features for reliable and stable readings across a wide range of applications.
This manual provides detailed instructions for the proper setup, operation, maintenance, and troubleshooting of your EC820 meter to ensure optimal performance and longevity.

Figure 1: Apera Instruments EC820 Benchtop Conductivity Meter Kit. The image shows the main meter unit, a flexible electrode arm and holder, and the conductivity electrode submerged in a beaker of liquid.
2. Saugos priemonės
- Always handle the meter and electrodes with care to prevent damage.
- Do not immerse the main meter unit in water. The unit is IP54 water-resistant, but not submersible.
- Naudokite tik pateiktą maitinimo adapterį.
- Keep calibration solutions and samples away from children and pets. Follow safety data sheets (SDS) for handling specific chemicals.
- Ensure the meter is placed on a stable, level surface during operation.
- Prieš valydami ar atlikdami techninę priežiūrą, atjunkite maitinimą.
3. Pakuotės turinys
Patikrinkite, ar jūsų pakuotėje yra visi toliau išvardyti daiktai:
- EC820 precizinis stalinis laidumo matuoklis
- 2401T-F Platinum Glass-body Conductivity Electrode
- Lankstus elektrodo laikiklis ir laikiklis
- USB kabelis
- PC-Link Software Flash Drive
- Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
- Kalibravimo sprendimai:
- 84 µS/cm (1 vial)
- 1413 µS/cm (1 vial)
- 12.88 mS/cm (1 vial)

Figure 2: The 2401T-F Platinum Glass-body Conductivity Electrode, showing its connection cables.
4. Sąranka
- Unpack the Meter: Atsargiai išimkite visus komponentus iš pakuotės.
- Assemble Electrode Arm: Attach the flexible electrode arm to the meter unit. Ensure it is securely fastened.
- Connect Electrode: Plug the 2401T-F Conductivity Electrode into the designated "COND" port on the back of the meter. Ensure the connection is firm.

Figure 3: Rear panel of the EC820 meter, highlighting the "COND" and "TEMP" ports, along with RS232 and DC 9V power input.
- Prijunkite maitinimą: Plug the power adapter into the "DC 9V" port on the back of the meter, then connect it to a standard 120V AC 60Hz power outlet.
- Pradinis įjungimas: Press the power button to turn on the meter. The LCD display should illuminate.
5. Naudojimo instrukcijos
5.1. Kalibravimas
The EC820 meter features 1 to 4 points auto-calibration. Regular calibration ensures accurate measurements.
- Paruoškite sprendimus: Use the provided 84 µS/cm, 1413 µS/cm, and 12.88 mS/cm calibration solutions. Ensure solutions are at room temperature.
- Skalavimo elektrodas: Rinse the conductivity electrode thoroughly with distilled or deionized water before and after each measurement or calibration. Gently blot dry with a lint-free tissue.
- Įjunkite kalibravimo režimą: With the meter on, press the CAL button. The display will show "CAL" and prompt for the first calibration point.
- Pirmojo taško kalibravimas: Immerse the electrode into the 1413 µS/cm standard solution. Wait for the reading to stabilize (indicated by a stable icon on the display). The meter will automatically recognize the solution. Press ENTER patvirtinti.
- Subsequent Points (Optional): For higher accuracy, proceed with 84 µS/cm and 12.88 mS/cm solutions following the same procedure. The meter will guide you through the process.
- Išeiti iš kalibravimo: After completing desired calibration points, press the PRIEMONĖS button to return to measurement mode.
Pastaba: The meter has a self-diagnosis function to help identify and fix common calibration problems. Refer to the on-screen prompts.
5.2. Matavimas
- Skalavimo elektrodas: Nuplaukite elektrodą distiliuotu vandeniu.
- Panardinamasis elektrodas: Panardinkite elektrodą įample solution, ensuring the sensor is fully submerged.
- Palaukite stabilumo: Allow the reading to stabilize. The stable reading icon will appear on the LCD.
- Įrašo skaitymas: Read the Conductivity (EC), TDS, Salinity, or Resistivity value displayed on the large LCD. Temperature will also be displayed simultaneously.
- Keisti vienetus: Paspauskite VIENETAS button to cycle through different measurement units (e.g., µS/cm, mS/cm, ppm, ppt).
- Duomenų saugykla: The meter supports GLP data management with 500 groups of data storage. Use the M+ / R-M buttons to store and recall data.
- Duomenų eksportavimas: Connect the meter to a PC via the USB cable and use the provided PC-Link software (Windows-based system only) to export data.
6. Priežiūra
- Elektrodų valymas:
- Regularly rinse the electrode with distilled or deionized water.
- For stubborn deposits, refer to the electrode's specific cleaning instructions. Generally, a mild detergent solution or a specific electrode cleaning solution can be used, followed by thorough rinsing.
- Do not scratch or damage the platinum black sensor.
- Elektrodų laikymas:
- Always store the conductivity electrode in its protective cap filled with a proper electrode storage solution (e.g., 3M KCl solution or Apera Instruments storage solution). Never store dry.
- Skaitiklio valymas: Wipe the main meter unit with a clean, damp audinys. Nenaudokite abrazyvinių valiklių ar tirpiklių.
- Programinės įrangos atnaujinimai: Check the Apera Instruments website periodically for any software updates for the PC-Link software.

4 pav. Iš arti view of the platinum glass-body conductivity electrode tip, which requires careful handling and proper storage.
7. Problemų sprendimas
| Problema | Galima priežastis | Sprendimas |
|---|---|---|
| Matuoklis neįsijungia | Nėra maitinimo šaltinio; sugedęs adapteris; atsilaisvinęs kontaktas. | Check power adapter connection and outlet. Ensure adapter is functional. |
| Nestabilūs rodmenys | Dirty electrode; air bubbles on sensor; incorrect calibration; temperature fluctuations. | Clean electrode; gently tap electrode to remove bubbles; recalibrate meter; ensure sample temperature is stable. |
| Kalibravimo klaida | Contaminated calibration solution; incorrect solution used; damaged electrode. | Use fresh calibration solutions; ensure correct solution for calibration point; inspect electrode for damage. |
| No data transfer to PC | Incorrect USB connection; software not installed or running; driver issues. | Verify USB cable connection; ensure PC-Link software is installed and open; check device manager for driver issues. |
8. Specifikacijos
| Parametras | Vertė |
|---|---|
| Modelio numeris | AI532 |
| Matavimo parametrai | Conductivity (EC), TDS, Salinity, Resistivity, Temperature |
| Laidumo diapazonas | 0 to 2,000 mS |
| Tikslumas | ±0.5 % FS |
| Kalibravimo taškai | 1 to 4 points auto-calibration (84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm) |
| Temperatūros kompensavimas | Automatinė temperatūros kompensacija (ATC) |
| Duomenų saugykla | 500 grupių (atitinka GLP reikalavimus) |
| Ryšys | USB for PC data export |
| Ekrano tipas | Didelis apšviestas LCD ekranas |
| Maitinimo šaltinis | 9V DC, 300mA (120V AC 60Hz adapter included) |
| Apsauga nuo patekimo | IP54 Atsparus vandeniui |
| Sertifikatai | CE, RoHS, ISO 9001:2015 |
| Pakuotės matmenys | 10.63 x 9.88 x 6.61 colio |
| Svoris | 6.5 svaro |
9. Garantija ir palaikymas
Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of quality and reliability. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments websvetainę.
For assistance with troubleshooting, calibration, or general inquiries, please contact Apera Instruments customer service. Contact details can typically be found on their official websvetainėje arba produkto pakuotėje.
Internetiniai ištekliai: Apera Instruments Store on Amazon