1.1 Produktas baigtasview

The Hach DR300 Pocket Colorimeter is a portable, waterproof instrument designed for precise measurement of Free and Total Chlorine in water samples. It provides reliable results for various water quality applications, including groundwater, surface water, and municipal water testing.

1.2 Pagrindinės savybės

  • Accurate Chlorine Measurement: Precisely detects Free Chlorine from 0.05–4.00 mg/L and Total Chlorine from 0.1–10.0 mg/L.
  • Rugged and Waterproof Design: Built to withstand harsh conditions with an IP67 rating, making it durable for field testing in tough environments.
  • Intuitive and Simple to Operate: Features a user-friendly interface and a large, backlit display for easy readings in any lighting.
  • Compatible with Hach Reagents: Designed to work seamlessly with EPA-approved Hach reagents and test methods to ensure valid and consistent results.
  • Portable and Reliable: Compact and lightweight for quick deployment, offering the dependability needed for professional water quality analysis.
Hach DR300 Pocket Colorimeter

Priekyje view of the Hach DR300 Pocket Colorimeter, displaying its screen and control buttons.

2. Kas įskaičiuota

The Hach DR300 Pocket Colorimeter kit typically includes the following components:

  • DR300 Pocket Colorimeter unit
  • Sturdy custom carrying case
  • 4 AAA batteries (pre-installed or included separately)
  • 2 glass sample ląstelės
  • 2 plastic sample ląstelės
  • Instrukcijų vadovas (šis dokumentas)

Note: Reagents for chlorine testing must be purchased separately.

Hach DR300 Pocket Colorimeter kit in carrying case

The Hach DR300 Pocket Colorimeter and its accessories, including sample cells and user manual, neatly organized within its custom carrying case.

3. Sąranka ir paruošimas

3.1 Akumuliatoriaus įdėjimas

The DR300 operates on 4 AAA batteries. Ensure they are correctly installed according to the polarity markings inside the battery compartment. The battery compartment is located on the rear of the device.

3.2 Pradinis įjungimas

Press the power button (usually marked with a power symbol) to turn on the device. The display should illuminate and show the initial measurement screen.

Hand holding Hach DR300 Pocket Colorimeter

A hand demonstrating the ergonomic design and portability of the Hach DR300 Pocket Colorimeter.

4. Operacija

4.1 General Measurement Procedure

The DR300 uses a photometric method for chlorine analysis. Always ensure sample cells are clean and free of scratches or fingerprints. Use Hach-approved reagents for accurate results.

4.2 Measuring Free Chlorine

  1. Užpildykite švarią sample cell with 10 mL of the water sample. This is your blank.
  2. Place the blank into the cell holder and close the lid.
  3. Press the zero button (often marked with a square or 'ZERO') to zero the instrument. The display will show "0.00".
  4. Remove the blank. Add the appropriate Hach DPD Free Chlorine reagent to a second 10 mL sample cell containing the water sample. Sumaišykite, kad sumaišytumėte.
  5. Wait for the specified reaction time (refer to reagent instructions, typically 1 minute).
  6. Place the reacted sample into the cell holder and close the lid.
  7. Press the read button (often marked with a checkmark or 'READ'). The instrument will display the Free Chlorine concentration in mg/L.

4.3 Measuring Total Chlorine

  1. Follow steps 1-3 from the Free Chlorine measurement to zero the instrument.
  2. Remove the blank. Add the appropriate Hach DPD Total Chlorine reagent to a second 10 mL sample cell containing the water sample. Sumaišykite, kad sumaišytumėte.
  3. Wait for the specified reaction time (refer to reagent instructions, typically 3 minutes).
  4. Place the reacted sample into the cell holder and close the lid.
  5. Press the read button. The instrument will display the Total Chlorine concentration in mg/L.
Technician using Hach DR300 Pocket Colorimeter in an industrial setting

A technician in a hard hat and safety glasses using the Hach DR300 Pocket Colorimeter to test water quality in an industrial environment, demonstrating its field applicability.

5. Priežiūra

5.1 Prietaiso valymas

Wipe the exterior of the colorimeter with a damp šluoste. Nenaudokite abrazyvinių valiklių ar tirpiklių. Įsitikinkite, kadample cell compartment is clean and dry before storage.

5.2 Cleaning SampLe Cells

Plauti sample cells thoroughly with distilled or deionized water after each use. Avoid scratching the glass or plastic surfaces, as this can affect measurement accuracy.

5.3 Saugykla

Store the DR300 and its accessories in the provided carrying case in a cool, dry place, away from direct sunlight and extreme temperatures.

6. Problemų sprendimas

ProblemaGalima priežastisSprendimas
Įrenginys neįsijungiaIšsikrovusios baterijosReplace batteries (4 AAA).
Ekrane rodomas pranešimas „Klaida“Sample cell dirty or scratched; Reagent expired or incorrect; Incorrect procedureŠvarus sample cell; Check reagent expiry and type; Review operation steps.
Netikslūs rodmenysSample cell not clean; Incorrect zeroing; Reagent issue; Temperature variationUžtikrinti sample cells are spotless; Re-zero the instrument; Use fresh, correct reagents; Allow samples to reach ambient temperature.
Ekranas blankus arba neįskaitomasLow batteries; Backlight settingReplace batteries; Adjust backlight setting if available.

7. Specifikacijos

  • Modelis: Cl, Free+Total, MR
  • Measurement Range (Free Chlorine): 0.05–4.00 mg/L Cl₂
  • Measurement Range (Total Chlorine): 0.1–10.0 mg/L Cl₂
  • Atsparumo vandeniui įvertinimas: IP67
  • Maitinimo šaltinis: 4 x AAA baterijos (pridedamos)
  • Matmenys (paketas): 9.09 x 6.93 x 5.75 colio
  • Svoris (pakuotės): 2.88 svaro
  • Gamintojas: Hach

8. Garantija ir palaikymas

For warranty information, technical support, or to purchase replacement parts and reagents, please contact Hach customer service or visit the official Hach website. Refer to the contact information provided with your product documentation or on the manufacturer's websvetainę.

You can visit the HACH Store on Amazon for more information: HACH Official Store

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.