1. Įvadas

This manual provides comprehensive instructions for the proper use and maintenance of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit. This kit is designed to ensure the optimal performance and longevity of your WTW CellOx Dissolved Oxygen (DO) probes.

The kit includes essential components for routine care, such as cleaning solution, abrasive foil, membrane heads, and electrolyte solution, all crucial for accurate and reliable dissolved oxygen measurements.

2. Rinkinio turinys

The WTW 202706 Heavy-Duty DO Probe Maintenance Kit contains the following items:

  • Cleaning solution for oxygen sensors (30 ml)
  • Electrolyte solution for oxygen sensors (50 ml)
  • Abrasive foil
  • Three membrane heads
Contents of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit, including two bottles of solution, a clear protective cap, a black probe component, and a yellow protective tube.

1 pav. Baigtaview of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit components. Visible are bottles of cleaning and electrolyte solutions, a protective cap, a probe component, and a yellow protective tube.

3. Sąranka ir paruošimas

Before performing maintenance, ensure you have a clean, dry workspace and any necessary personal protective equipment (PPE).

  1. Identify Probe Type: Confirm that your DO probe is a WTW CellOx probe, as this kit is specifically designed for these models.
  2. Surinkite komponentus: Lay out all components from the maintenance kit and your DO probe.
  3. Skaitykite instrukcijas: Familiarize yourself with the entire maintenance procedure before beginning.

4. Operating Procedures (Probe Maintenance)

Regular maintenance is crucial for accurate DO measurements. Follow these steps for proper probe care:

4.1. Zondo valymas

  1. Rinse Probe: Gently rinse the probe with distilled water to remove any loose debris.
  2. Taikyti valymo tirpalą: Apply a small amount of the provided cleaning solution to a soft cloth or directly to the probe's membrane area.
  3. Švelniai valykite: Carefully wipe the membrane and probe body. For stubborn deposits, use the abrasive foil provided, but apply minimal pressure to avoid damaging the membrane.
  4. Kruopščiai nuplaukite: Rinse the probe thoroughly with distilled water after cleaning to remove all traces of the cleaning solution.

4.2. Replacing the Membrane Head

The membrane head should be replaced periodically or if damaged.

  1. Unscrew Old Head: Carefully unscrew the old membrane head from the probe body.
  2. Prepare New Head: Take one of the new membrane heads from the kit.
  3. Fill with Electrolyte: Fill the new membrane head with the provided electrolyte solution. Ensure there are no air bubbles trapped inside. Gently tap the head to dislodge any bubbles.
  4. Screw On New Head: Carefully screw the new membrane head onto the probe body. Ensure it is finger-tight to prevent leaks, but do not overtighten.

4.3. Electrolyte Replacement

The electrolyte solution should be replaced when replacing the membrane head or if the probe's response becomes sluggish.

  1. Drain Old Electrolyte: If not replacing the membrane head, carefully drain the old electrolyte solution from the existing membrane head.
  2. Refill with Fresh Electrolyte: Fill the membrane head with fresh electrolyte solution from the kit, ensuring no air bubbles are present.

5. Sandėliavimas ir tvarkymas

Proper storage of the kit components and the probe itself ensures their longevity and readiness for use.

  • Store the kit components in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Keep bottles of cleaning and electrolyte solutions tightly sealed when not in use.
  • Always store the DO probe with its protective cap filled with a small amount of electrolyte solution or in a suitable storage solution as recommended by WTW.

6. Įprastų problemų šalinimas

If you encounter issues with your DO probe after maintenance, consider the following:

ProblemaGalima priežastisSprendimas
Nestabilūs rodmenysAir bubbles in membrane head, dirty membrane, depleted electrolyte.Tap membrane head to remove bubbles, clean membrane, replace electrolyte.
Lėtas reakcijos laikasOld or contaminated electrolyte, damaged membrane.Replace electrolyte and/or membrane head.
Jokio skaitymoCompletely dry membrane, severe probe damage.Refill with electrolyte and re-condition probe. If still no reading, probe may need replacement.

7. Specifikacijos

Key specifications for the WTW 202706 Heavy-Duty DO Probe Maintenance Kit:

  • Modelio numeris: 202706
  • Suderinamumas: Designed for WTW CellOx DO probes
  • Komplekto turinys: Cleaning solution, abrasive foil, 3 membrane heads, electrolyte solution
  • Pakuotės matmenys: 8.79 x 5.49 x 2.64 colio
  • Pakuotės svoris: 9.44 uncijos
  • Gamintojas: COLE-PARMER (for distribution)
  • Prekinis ženklas: WTW

8. Garantija ir palaikymas

The WTW 202706 Heavy-Duty DO Probe Maintenance Kit comes with a 1 metų garantija nuo pirkimo datos.

For technical support, warranty claims, or further assistance, please contact your authorized WTW distributor or visit the official WTW websvetainėje. Kreipdamiesi pagalbos, būtinai turėkite po ranka savo produkto modelio numerį (202706) ir pirkimo įrodymą.