Hantek DSO2D50

Hantek DSO2D50 Digital Storage Oscilloscope User Manual

Model: DSO2D50

1. Įvadas

The Hantek DSO2D50 is a high-performance digital storage oscilloscope designed for precise signal analysis and measurement. It features a 500MHz bandwidth, 2GSa/s real-time sampling rate, and an 80K memory depth. This instrument also integrates a 25MHz arbitrary waveform generator, making it a versatile tool for various electronic testing and development applications.

This manual provides essential information for the safe and effective operation of your DSO2D50 oscilloscope. Please read it thoroughly before use and retain it for future reference.

Hantek DSO2D50 Digital Storage Oscilloscope front view

1.1 pav.: priekis view of the Hantek DSO2D50 Digital Storage Oscilloscope.

Pagrindinės funkcijos:

2. Saugos informacija

ĮSPĖJIMAS:

Rear panel of Hantek DSO2D50 with warning label

Figure 2.1: Rear panel showing power input and safety warning.

Pavojingas ttage inside, do not remove the cover unless by specified personnel.

3. Pakuotės turinys

Patikrinkite, ar jūsų pakuotėje yra šie elementai:

If any items are missing or damaged, please contact your dealer or Hantek customer support immediately.

4. Produktas baigtasview

4.1 Front Panel Controls and Connectors

The front panel of the DSO2D50 features the display, control buttons, knobs, and input connectors for channels and the arbitrary waveform generator output.

Išsamus view of Hantek DSO2D50 control panel

4.1 pav. Išsamus view of the control panel and input connectors.

Key areas include:

4.2 galinio skydelio jungtys

The rear panel includes the power input and ventilation.

5. Sąranka

5.1 Maitinimo jungtis

  1. Įsitikinkite, kad osciloskopo maitinimo jungiklis yra IŠJUNGTOJE padėtyje.
  2. Prijunkite pateiktą maitinimo laidą prie kintamosios srovės maitinimo įvesties, esančios osciloskopo galiniame skydelyje.
  3. Kitą maitinimo laido galą prijunkite prie įžeminto kintamosios srovės lizdo.

5.2 Zondo prijungimas ir kompensavimas

  1. Connect a passive probe to one of the CH1 or CH2 BNC input connectors.
  2. Set the probe attenuation switch (if applicable) to 10X.
  3. Connect the probe ground clip to the oscilloscope's ground terminal (usually a dedicated terminal or the ground ring of the BNC connector).
  4. Connect the probe tip to the probe compensation output (usually a square wave test signal output on the front panel).
  5. Įjunkite osciloskopą.
  6. Paspauskite AUTOMATINIS NUSTATYMAS button. The oscilloscope will automatically adjust settings to display the compensation signal.
  7. Adjust the compensation screw on the probe until a flat-top square wave is displayed on the screen. This ensures accurate measurements.

6. Naudojimo instrukcijos

6.1 Pagrindinis bangos formos ekranas

  1. Įjunkite osciloskopą.
  2. Connect a compensated probe to CH1.
  3. Connect the probe tip to the signal source you wish to measure.
  4. Paspauskite AUTOMATINIS NUSTATYMAS button. The oscilloscope will automatically adjust the vertical, horizontal, and trigger settings to display a stable waveform.
  5. Naudokite VOLTS/DIV knob to adjust the vertical scale (ampšviesa).
  6. Naudokite SEC/DIV knob to adjust the horizontal scale (time base).
  7. Naudokite POZICIJA rankenėlės, skirtos bangos formai perkelti vertikaliai arba horizontaliai.

6.2 1MΩ/50Ω Impedance Switching

The DSO2D50 supports switching between 1MΩ (high impedance) and 50Ω (low impedance) input impedance. This is crucial for adapting to different measurement scenarios and ensuring signal integrity, especially for high-frequency signals or when matching transmission lines.

Oscilloscope screen showing impedance switching menu

Figure 6.1: Menu for selecting input impedance.

To switch impedance:

  1. Paspauskite CH1 MENU or CH2 MENU button for the desired channel.
  2. Navigate through the menu options using the function buttons (F0-F6) until you find the "Impedance" setting.
  3. Pasirinkite bet kurį 1MΩ or 50Ω as required for your measurement.
  4. Pastaba: When using 50Ω impedance, ensure your signal source is also 50Ω terminated to prevent reflections.

6.3 Measurement and Statistical Functions

The oscilloscope provides 32 automatic measurement functions and threshold measurement functions. These tools support statistical analysis, allowing you to display five measurement values: current value, average value, maximum value, minimum value, and standard deviation.

Oscilloscope screen displaying multiple automatic measurements

6.2 pav. Pvzample of automatic measurement display.

To access measurement functions:

  1. Paspauskite PAMATUOKITE mygtuką.
  2. Use the function buttons to select the desired measurement type (e.g., Frequency, Period, Vpp, Vmax, Vmin, Rise Time, Fall Time).
  3. The selected measurements will appear on the screen.
  4. Į view statistical values, navigate to the statistical menu within the MEASURE options.

6.4 Built-in Arbitrary Waveform Generator (AWG)

The DSO2D50 includes a 25MHz arbitrary waveform generator capable of outputting five standard waveforms: sine wave, square wave, triangle wave, exponential wave, noise, and DC. It also supports custom arbitrary waveform output.

Oscilloscope screen showing arbitrary waveform generator settings

Figure 6.3: Arbitrary Waveform Generator settings interface.

To use the AWG:

  1. Connect a BNC cable from the EXT TRIG/GEN OUT connector to the input of the device under test or an oscilloscope channel.
  2. Paspauskite WAVE GEN mygtuką.
  3. Use the function buttons to select the desired waveform type (Sine, Square, Triangle, etc.).
  4. Koreguokite parametrus, tokius kaip dažnis, Amplitude, and Offset using the corresponding menu options and knobs.
  5. Ensure the AWG output is enabled within the menu.

6.5 Suveikimo režimai

The oscilloscope offers 9 triggering modes to stabilize and capture specific events in your waveform. Common modes include Edge, Pulse, Video, Slope, and more advanced options for serial bus decoding.

To set up a trigger:

  1. Paspauskite TRIG MENU mygtuką.
  2. Select the desired trigger type (e.g., Edge, Pulse).
  3. Adjust the trigger source (CH1, CH2, EXT), slope (rising/falling edge), and level using the TRIGGER LEVEL knob and menu options.
  4. The trigger level is indicated by a horizontal line on the display.

6.6 Duomenų saugojimas ir atkūrimas

The DSO2D50 allows you to save and recall various data types, including CSV files, images, reference waveforms, and waveform data, to a USB storage device.

To save data:

  1. Insert a USB flash drive into the USB Host port on the front panel.
  2. Paspauskite SAVE TO USB mygtuką.
  3. Select the desired save type (e.g., Image, Waveform, CSV) and file format from the menu.
  4. Patvirtinkite išsaugojimo operaciją.

6.7 Išplėstinės funkcijos

7. Priežiūra

7.1 Valymas

7.2 Saugykla

7.3 Kalibravimas

The DSO2D50 is factory calibrated. For continued accuracy, periodic calibration by qualified service personnel is recommended, typically every 1-2 years depending on usage.

8. Problemų sprendimas

Problema Galima priežastis Sprendimas
Po įjungimo ekrano nėra. Power cord not connected, power switch off, or power supply issue. Check power cord connection. Ensure power switch is ON. Verify power outlet functionality.
Nerodoma bangos forma. Probe not connected, signal source off, vertical/horizontal scale incorrect, or trigger not set. Check probe connection. Ensure signal source is active. Press AUTOMATINIS NUSTATYMAS. Adjust VOLTS/DIV and SEC/DIV. Check trigger settings.
Nestabili bangos forma. Incorrect trigger settings. Adjust trigger level. Change trigger type (e.g., from Auto to Normal). Check trigger source.
Matavimai netikslūs. Probe compensation incorrect, probe attenuation setting mismatch, or incorrect impedance. Perform probe compensation. Ensure probe attenuation (1X/10X) matches oscilloscope setting. Verify input impedance (1MΩ/50Ω).
USB storage not recognized. Incompatible USB drive, drive not formatted correctly, or drive issue. Try a different USB drive. Ensure drive is formatted to FAT32.

9. Specifikacijos

The following table outlines the key specifications for the Hantek DSO2D50 Digital Storage Oscilloscope.

Parametras Specifikacija
Modelis DSO2D50
Kanalai 2 Analog Channels + 1 External Trigger Channel
Pralaidumas 500 MHz
Samplingo norma 2 GSa/s (Real-time)
Atminties gylis 80 Kpts
Savavališkas bangų formų generatorius 25 MHz (Sine, Square, Triangle, Exponential, Noise, DC)
Vertikalus jautrumas 500 µV/div - 10 V/div
Įvesties varža 1 MΩ ±2% || 15 pF ±5 pF; 50 Ω ±2%
Maksimali įvesties apimtistage 300 V CAT II (1MΩ)
Ekranas 7 colių TFT LCD
Suaktyvinimo režimai Edge, Pulse, Video, Slope, Overtime, Window, Pattern, Interval, UART, LIN, CAN, SPI, IIC
Automatiniai matavimai 32 types, with statistical analysis
Sąsaja USB priegloba, USB įrenginys
Maitinimo šaltinis 100-240V kintamoji srovė, 50/60Hz
Energijos suvartojimas < 50W
Matmenys (Refer to package dimensions, typically 5.91 x 3.94 x 0.79 inches for package, actual unit size will be larger)
Svoris 12.1 svaro (prekės svoris)

Pastaba: Specifications are subject to change without notice. For the most current specifications, please refer to the manufacturer's official websvetainę.

9.1 Modelių palyginimas

The Hantek DSO2D50 is part of a series of oscilloscopes. The following table provides a comparison with other models in the series, highlighting key differences.

Table comparing Hantek oscilloscope models

Figure 9.1: Model comparison table for Hantek oscilloscopes.

This table helps to understand the DSO2D50's position within the product line, specifically noting its 500MHz bandwidth, 2GSa/s sampling rate, 80K storage depth, and integrated arbitrary function generator (AFG).

10. Garantija ir palaikymas

10.1 Informacija apie garantiją

Hantek products are typically covered by a limited warranty against defects in materials and workmanship. The specific warranty period and terms may vary by region and product. Please refer to the warranty card included with your product or visit the official Hantek websvetainėje rasite išsamią informaciją apie garantiją.

Išsaugokite pirkimo kvitą kaip pirkimo įrodymą garantiniams reikalavimams.

10.2 Pagalba klientams

For technical assistance, troubleshooting, or service inquiries, please contact Hantek customer support:

When contacting support, please have your product model (DSO2D50) and serial number ready.

Susiję dokumentai - DSO2D50

Preview Hantek DSO2D20 Series Digital Oscilloscope User Manual
Comprehensive user manual for the Hantek DSO2D20 series digital oscilloscopes, detailing features, operation, troubleshooting, and specifications for accurate signal analysis.
Preview Hantek DPO7000 Series Digital Oscilloscope Datasheet
Comprehensive datasheet for the Hantek DPO7000 Series Digital Oscilloscope, detailing features, technical specifications, ordering information, and optional upgrades for these versatile 8-in-1 instruments.
Preview Uživatelská příručka Hantek HDG3000B: Generátor signálu s libovolnou vlnou
Podrobná uživatelská příručka pro generátor signálu Hantek řady HDG3000B. Obsahuje informace o bezpečnosti, specifikacích produktu, rychlém startu, nastavení, výstupu vlnových forem, modulaci, sweepu, burstu, čítači, ukládání a dálkovém ovládání. Tato příručka je navržena tak, aby uživatelům pomohla efektivně využívat pokročilé funkce generátoru pro jejich testovací a měřicí potřeby.
Preview „Hantek2D82“ automobilių diagnostikos osciloskopo naudotojo vadovas | Išsamus vadovas
Susipažinkite su „Hantek2D82“ automobilių diagnostikos osciloskopu. Šiame naudotojo vadove pateikiama išsami informacija apie jo funkcijas, veikimą, saugą, technines specifikacijas ir trikčių šalinimą, skirtą automobilių diagnostikai, DMM ir bangų formos generavimui.
Preview Hantek DSO2D20 Series Digital Oscilloscope Quick Guide
A concise and SEO-optimized HTML guide for the Hantek DSO2D20 Series Digital Oscilloscope, covering safety, product overview, preparation, and remote control.
Preview Hantek DSO3062L/DSO3062AL Digital Oscilloscope User Manual
Explore the Hantek DSO3062L/DSO3062AL Digital Oscilloscope User Manual. This guide provides detailed instructions on setup, operation, safety precautions, advanced functions, and application examples for accurate electronic measurements.